Title

Prevention of airway fires: testing the safety of endotracheal tubes and surgical devices in a mechanical model

Publication Date

2015

Journal Title

Am J Otolaryngol

Abstract

OBJECTIVES: This study was designed to assess the ability of carbon dioxide (CO2) lasers and radiofrequency ablation devices (Coblator) (ArthoCare Corporation, Sunnyvale, CA) to ignite either a non-reinforced (polyvinylchloride) endotracheal tube (ETT) or an aluminum and fluoroplastic wrapped silicon ("laser safe") ETT at varying titrations of oxygen in a mechanical model of airway surgery. METHODS: Non-reinforced and laser safe ETTs were suspended in a mechanical model imitating endoscopic airway surgery. A CO2 laser set at 5-30 watts was fired at the ETT at oxygen concentrations ranging from 21% to 88%. The process was repeated using a radiofrequency ablation (RFA) device. All trials were repeated to ensure accuracy. RESULTS: The CO2 laser ignited a fire when contacting a non-reinforced ETT in under 2 seconds at oxygen concentrations as low as 44%. The CO2 laser could not ignite a laser safe ETT under any conditions, unless it struck the non-reinforced distal tip of the ETT. With the RFA, a fire could not be ignited with either reinforced or non-reinforced ETTs. CONCLUSIONS: RFA presents no risk of ignition in simulated airway surgery. CO2 lasers should be utilized with a reinforced ETT or no ETT, as fires can easily ignite when lasers strike a non-reinforced ETT. Decreasing the fraction of inspired oxygen reduces the risk of fire.

Volume Number

36

Issue Number

1

Pages

63-6

Document Type

Article

EPub Date

2014/12/03

Status

Faculty

Facility

School of Medicine

Primary Department

Otolaryngology

PMID

25456513

DOI

10.1016/j.amjoto.2014.10.009

For the public and Northwell Health campuses

Share

COinS