Multicenter validation of the VITEK MS v2.0 MALDI-TOF mass spectrometry system for the identification of fastidious gram-negative bacteria
Publication Date
2014
Journal Title
Diagn Microbiol Infect Dis
Abstract
The VITEK MS v2.0 MALDI-TOF mass spectrometry system's performance in identifying fastidious gram-negative bacteria was evaluated in a multicenter study. Compared with the reference method (DNA sequencing), the VITEK MS system provided an accurate, species-level identification for 96% of 226 isolates; an additional 1% were accurately identified to the genus level. (C) 2014 Elsevier Inc. All rights reserved.
Volume Number
78
Issue Number
2
Pages
129-131
Document Type
Article
Status
Faculty, Northwell Researcher
Facility
School of Medicine; Northwell Health
Primary Department
Pathology and Laboratory Medicine
Additional Departments
Molecular Medicine
PMID
DOI
10.1016/j.diagmicrobio.2013.08.013