Rotation effects on the target-volume margin determination

Publication Date

2015

Journal Title

Phys Med

Abstract

Rotational setup errors are usually neglected in most clinical centers. An analytical formula is developed to determine the extra margin between clinical target volume (CTV) and planning target volume (PTV) to account for setup errors. The proposed formula corrects for both translational and rotational setup errors and then incorporated into margin determination for PTV. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Volume Number

31

Issue Number

1

Pages

80-84

Document Type

Article

Status

Faculty

Facility

School of Medicine

Primary Department

Radiation Medicine

PMID

25455438

DOI

10.1016/j.ejmp.2014.10.076

For the public and Northwell Health campuses

Share