Publication Date

2015

Journal Title

J Appl Clin Med Phys

Abstract

The purpose of this study was to evaluate the use of diverging-cut aperture to minimize collimator contamination in proton therapy. Two sets of apertures with nondivergent and divergent edge were fabricated to produce a 10 cm x 10 cm field at the radiation isocenter of a single-room proton therapy unit. Transverse profiles were acquired in a scanning water tank with both aperture sets. Up to 9.5% extra dose was observed from aperture scattering near the field edges with the nondivergent aperture set at 2 cm above the water surface and remained 3.0% at depth of 10 cm. For the divergent set, the contamination was reduced to less than 3.5% and 1.3%, respectively. Our study demonstrated that scattering from apertures contaminated the dose distribution near the field edge at shallow depth. A diverging-cut aperture was capable of reducing the contamination and is recommended for use in passive scattering proton therapy, especially when critical organs are lateral and proximal to the target at shallow depth.

Volume Number

16

Issue Number

5

Pages

367-372

Document Type

Article

Status

Faculty

Facility

School of Medicine

Primary Department

Radiation Medicine

PMID

26699327

DOI

10.1120/jacmp.v16i5.5675


Share